Remote Commands
R&S
®
FPL1000
780
User Manual 1178.3370.02 ─ 11
Manual operation:
See
INPut<ip>:IMPedance
<Impedance>
This command selects the nominal input impedance of the RF input. In some applica-
tions, only 50
Ω
are supported.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Impedance>
50 | 75
numeric value
User-defined impedance from 50 Ohm to 100000000 Ohm
(=100 MOhm)
User-defined values are only available for the Spectrum applica-
tion, the I/Q Analyzer, and some optional applications.
*RST:
50 Ω
Default unit: OHM
Example:
INP:IMP 75
Manual operation:
See
See
INPut<ip>:SELect
<Source>
This command selects the signal source for measurements, i.e. it defines which con-
nector is used to input data to the R&S
FPL1000.
If no additional input options are installed, only RF input is supported.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Source>
RF
Radio Frequency ("RF INPUT" connector)
*RST:
RF
Manual operation:
See
INPut<ip>:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the
AUX PORT
connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S
FPL1000.
See the R&S
FPL1000 Getting Started manual for details.
Setting Basic Spectrum RF Measurement Parameters