![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 687](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987687.webp)
Remote Commands
R&S
®
FPL1000
687
User Manual 1178.3370.02 ─ 11
Suffix:
<n>
.
Parameters:
<Margin>
Range:
-200 to 200
Default unit: dB
Example:
CALC:PSE:MARG 100
Sets the threshold to 100 dB.
Manual operation:
See
CALCulate<n>:PSEarch:PSHow
<State>
CALCulate<n>:PEAKsearch:PSHow
<State>
This command turns the peak labels in the diagram on and off.
Peak labels are blue squares.
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:PSE:PSH ON
Marks all peaks with blue squares.
Manual operation:
See
CALCulate<n>:PSEarch:SUBRanges
<NumberPeaks>
CALCulate<n>:PEAKsearch:SUBRanges
<NumberPeaks>
This command defines the number of peaks included in the peak list.
After this number of peaks has been found, the R&S
FPL1000 stops the peak search
and continues the search in the next measurement range.
Suffix:
<n>
.
Parameters:
<NumberPeaks>
Range:
1 to 50
*RST:
25
Example:
CALC:PSE:SUBR 10
Sets 10 peaks per range to be stored in the list.
Manual operation:
See
Configuring and Performing Measurements