Remote Commands
R&S
®
FPL1000
827
User Manual 1178.3370.02 ─ 11
ALL
Selects all active traces and result tables (e.g. Result Summary,
marker peak list etc.) in the current application for export to an
ASCII file.
The <trace> parameter for the
command is ignored.
*RST:
SINGle
Manual operation:
See
"Export all Traces and all Table Results"
FORMat:DEXPort:XDIStrib
<XDistribution>
Defines how the x-values of the trace are determined in the frequency domain.
Parameters:
<XDistribution>
STARtstop | BINCentered
BINCentered
The full measurement span is divided by the number of sweep
points to obtain
bin
s. The x-value of the sweep point is defined
as the x-value at the center of the bin (bin/2).
STARtstop
(Default): The x-value of the first sweep point corresponds to the
starting point of the full measurement span. The x-value of the
last sweep point corresponds to the end point of the full mea-
surement span. All other sweep points are divided evenly
between the first and last points.
Example:
FORM:DEXP:XDIS BINC
Manual operation:
See
FORMat:DIMPort:TRACes
<Selection>
This command selects the data to be included in a data import file (see
For details on importing data see
Chapter 6.13.3, "How to Import Traces"
Parameters:
<Selection>
SINGle | ALL
SINGle
Only a single trace is selected for import, namely the one speci-
fied by the
mand.
Setting Basic Spectrum RF Measurement Parameters