Remote Commands
R&S
®
FPL1000
679
User Manual 1178.3370.02 ─ 11
[SENSe:]LIST:RANGe<ri>:BREak
<State>
This command controls the sweep for all ranges.
Suffix:
<ri>
.
irrelevant
Parameters:
<State>
ON | 1
The R&S
FPL1000 stops after measuring one range, and the
status bit number 10 in the
STAT:OPER
register is set.
To continue with the next range, use
.
OFF | 0
The R&S
FPL1000 sweeps all ranges in one go.
*RST:
0
Example:
LIST:RANG2:BRE ON
[SENSe:]LIST:RANGe<ri>:COUNt?
This command queries the number of ranges in the sweep list.
Suffix:
<ri>
.
irrelevant
Return values:
<Ranges>
Number of ranges in the sweep list.
Example:
LIST:RANG:COUN?
Usage:
Query only
[SENSe:]LIST:RANGe<ri>:DELete
This command removes a range from the sweep list.
Note that
●
you cannot delete the reference range
●
a minimum of three ranges is mandatory.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Example:
LIST:RANG2:DEL
[SENSe:]LIST:RANGe<ri>:DETector
<Detector>
This command selects the detector for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.