The Spectrum Application (RF Measurements)
R&S
®
FPL1000
99
User Manual 1178.3370.02 ─ 11
6. To start the measurement, select one of the following:
● [RUN SINGLE] key
● "Single Sweep" softkey in the "Sweep" menu
The defined number of sweeps are performed, then the measurement is stopped.
While the measurement is running, the [RUN SINGLE] key is highlighted. To abort
the measurement, press the [RUN SINGLE] key again. The key is no longer high-
lighted. The results are not deleted until a new measurement is started.
7. To repeat the same number of sweeps without deleting the last trace, select the
"Continue Single Sweep" softkey in the "Sweep" menu.
To start continuous sweeping
1. If you want to average the trace or search for a maximum over more (or less) than
10 sweeps, configure the "Sweep/Average Count" ("Sweep Config" dialog box, see
2. To start the measurement, select one of the following:
● [RUN CONT] key
● "Continuous Sweep" softkey in the "Sweep" menu
After each sweep is completed, a new one is started automatically. While the mea-
surement is running, the [RUN CONT] key is highlighted. To stop the measure-
ment, press the [RUN CONT] key again. The key is no longer highlighted. The
results are not deleted until a new measurement is started.
6.2.1.3
Measurement Examples - Measuring a Sinusoidal Signal
One of the most common measurement tasks that can be handled using a signal ana-
lyzer is determining the level and frequency of a signal. When measuring an unknown
signal, you can usually start with the presettings.
High input values
If levels higher than +30 dBm (=1
W) are expected or are possible, a power attenuator
must be inserted before the RF input of the analyzer. Otherwise, signal levels exceed-
ing 30
dBm can damage the RF attenuator or the input mixer. The total power of all
occurring signals must be taken into account.
Test setup
●
Connect the RF output of the signal generator to the RF input of the
R&S
FPL1000.
Measurements and Results