Remote Commands
R&S
®
FPL1000
611
User Manual 1178.3370.02 ─ 11
Note:
In order to synchronize to the end of a measurement sequence using *OPC,
*OPC? or *WAI you must use
SINGle
Sequence mode.
Parameters:
<Mode>
SINGle
Each measurement group is started one after the other in the
order of definition. All measurement channels in a group are
started simultaneously and performed once. After
all
measure-
ments are completed, the next group is started. After the last
group, the measurement sequence is finished.
CONTinuous
Each measurement group is started one after the other in the
order of definition. All measurement channels in a group are
started simultaneously and performed once. After
all
measure-
ments are completed, the next group is started. After the last
group, the measurement sequence restarts with the first one and
continues until it is stopped explicitely.
*RST:
CONTinuous
Manual operation:
See
SYSTem:SEQuencer
<State>
This command turns the Sequencer on and off. The Sequencer must be active before
any other Sequencer commands (
INIT:SEQ...
) are executed, otherwise an error will
occur.
A detailed programming example is provided in
Chapter 10.5.3, "Programming Exam-
ple: Performing a Sequence of Measurements"
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
The Sequencer is activated and a sequential measurement is
started immediately.
OFF | 0
The Sequencer is deactivated. Any running sequential measure-
ments are stopped. Further Sequencer commands
(
INIT:SEQ...
) are not available.
*RST:
0
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single Sequencer mode so each active measurement will
be performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
SYST:SEQ OFF
Manual operation:
See