Remote Commands
R&S
®
FPL1000
752
User Manual 1178.3370.02 ─ 11
Parameters:
<Factor>
1 to 100 PCT
*RST:
10
Default unit: PCT
Example:
//Couple frequency step size to span and define a step size fac-
tor
FREQ:CENT:STEP:LINK SPAN
FREQ:CENT:STEP:LINK:FACT 20PCT
Manual operation:
See
[SENSe:]FREQuency:OFFSet
<Offset>
This command defines a frequency offset.
If this value is not 0
Hz, the application assumes that the input signal was frequency
shifted outside the application. All results of type "frequency" will be corrected for this
shift numerically by the application.
See also
Parameters:
<Offset>
Range:
-1 THz to 1 THz
*RST:
0 Hz
Default unit: HZ
Example:
FREQ:OFFS 1GHZ
Manual operation:
See
[SENSe:]FREQuency:SPAN
<Span>
This command defines the frequency span.
If you set a span of 0
Hz in the Spectrum application, the R&S
FPL1000 starts a mea-
surement in the time domain.
Parameters:
<Span>
The minimum span for measurements in the frequency domain
is 10
Hz.
For SEM and spurious emission measurements, the minimum
span is 20
Hz.
Range:
0 Hz to fmax
*RST:
Full span
Default unit: Hz
Manual operation:
See
See
See
See
Setting Basic Spectrum RF Measurement Parameters