Remote Commands
R&S
®
FPL1000
754
User Manual 1178.3370.02 ─ 11
This includes:
●
Center frequency
●
Reference level
Example:
ADJ:ALL
Manual operation:
See
"Adjusting all Determinable Settings Automatically (Auto
[SENSe:]ADJust:CONFigure:LEVel:DURation
<Duration>
In order to determine the ideal reference level, the R&S
FPL1000 performs a measure-
ment on the current input data. This command defines the length of the measurement
if
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
MANual
.
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation:
See
"Changing the Automatic Measurement Time (Meastime
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
<Mode>
In order to determine the ideal reference level, the R&S
FPL1000 performs a measure-
ment on the current input data. This command selects the way the R&S
FPL1000
determines the length of the measurement .
Parameters:
<Mode>
AUTO
The R&S
FPL1000 determines the measurement length auto-
matically according to the current input data.
MANual
The R&S
FPL1000 uses the measurement length defined by
[SENSe:]ADJust:CONFigure:LEVel:DURation
on page 754.
*RST:
AUTO
Manual operation:
See
"Resetting the Automatic Measurement Time (Meastime
See
"Changing the Automatic Measurement Time (Meastime
Setting Basic Spectrum RF Measurement Parameters