Network Operation and Remote Control
R&S
®
FPL1000
564
User Manual 1178.3370.02 ─ 11
Bit No.
Meaning
2 to 14
Unused
15
This bit is always 0.
9.2.3
Reset Values of the Status Reporting System
The following table contains the different commands and events causing the status
reporting system to be reset. None of the commands, except
*RST
and
SYSTem:PRESet
, influence the functional instrument settings. In particular,
DCL
does
not change the instrument settings.
Table 9-15: Resetting the status reporting system
Event
Switching on supply
voltage
Power-On-Status-
Clear
DCL,
SDC
(Device
Clear,
Selected
Device
Clear)
*RST or
SYS-
Tem:PRE
Set
STA-
Tus:PRE-
Set
*CLS
Effect
0
1
Clear STB, ESR
-
yes
-
-
-
yes
Clear SRE, ESE
-
yes
-
-
-
-
Clear PPE
-
yes
-
-
-
-
Clear EVENt parts of the regis-
ters
-
yes
-
-
-
yes
Clear ENABle parts of all
OPERation and QUEStionable
registers;
Fill ENABle parts of all other
registers with "1".
-
yes
-
-
yes
-
Fill PTRansition parts with "1";
Clear NTRansition parts
-
yes
-
-
yes
-
Clear error queue
yes
yes
-
-
-
yes
Clear output buffer
yes
yes
yes
1)
1)
1)
Clear command processing
and input buffer
yes
yes
yes
-
-
-
1) The first command in a command line that immediately follows a <PROGRAM MESSAGE TERMINA-
TOR> clears the output buffer.
9.3
GPIB Languages
The R&S
FPL1000 analyzer family supports a subset of the GPIB commands used by
other devices. Thus it can emulate other devices in order to use existing remote control
programs.
GPIB Languages