![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 135](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987135.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
135
User Manual 1178.3370.02 ─ 11
Standard
The main measurement settings can be stored as a standard file. When such a stan-
dard is loaded, the required channel and general measurement settings are automati-
cally set on the R&S
FPL1000. However, the settings can be changed. Predefined
standards are available for standard measurements, but standard files with user-
defined configurations can also be created.
Predefined Standards ← Standard
Predefined standards contain the main measurement settings for standard measure-
ments. When such a standard is loaded, the required channel settings are automati-
cally set on the R&S
FPL1000. However, you can change the settings.
The predefined standards contain the following settings:
●
Channel bandwidths
●
Channel spacings
●
Detector
●
Trace Average setting
●
Resolution Bandwidth (RBW)
●
Weighting Filter
For details on the available standards, see
Chapter 6.2.4.8, "Reference: Predefined
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet
User Standards ← Standard
Access
: "CP / ACLR Config" > "General Settings" tab > "Manage User Standards"
In addition to the predefined standards, you can save your own standards with your
specific measurement settings in an XML file so you can use them again later. User-
defined standards are stored on the instrument in the
C:\ProgramData\Rohde-Schwarz\ZNL-FPL\acp_std
directory.
Note:
User standards created on an analyzer of the R&S FSP family are compatible to
the R&S
FPL1000. User standards created on an R&S
FPL1000, however, are not
necessarily compatible to the analyzers of the R&S FSP family and may not work
there.
The following parameter definitions are saved in a user-defined standard:
●
Number of adjacent channels
●
Channel bandwidth of transmission (Tx), adjacent (Adj) and alternate (Alt) chan-
nels
●
Channel spacings
●
Weighting filters
●
Resolution bandwidth
●
Video bandwidth
●
Detector
●
ACLR limits and their state
●
"Sweep Time" and "Sweep Time" coupling
●
Trace and power mode
Save the current measurement settings as a user-defined standard, load a stored mea-
surement configuration, or delete an existing configuration file.
Measurements and Results