![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 670](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987670.webp)
Remote Commands
R&S
®
FPL1000
670
User Manual 1178.3370.02 ─ 11
●
You can only use power classes for which limits are defined.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1 to 3
<pc>
1..n
power class
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:LIM:ESP:PCL1 ON
Activates the first defined power class.
Manual operation:
See
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:LIMit[:STATe]
<State>
This command selects the limit check mode for each power class.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1 to 3
<pc>
1..n
power class
Parameters:
<State>
ABSolute | RELative | AND | OR
ABSolute
Evaluates only limit lines with absolute power values
RELative
Evaluates only limit lines with relative power values
AND
Evaluates limit lines with relative and absolute power values. A
negative result is returned if both limits fail.
OR
Evaluates limit lines with relative and absolute power values. A
negative result is returned if at least one limit failed.
*RST:
REL
Example:
CALC:LIM:ESP:PCL:LIM ABS
Manual operation:
See
Configuring and Performing Measurements