Remote Commands
R&S
®
FPL1000
684
User Manual 1178.3370.02 ─ 11
Parameters:
<Points>
For more information on sweep points see
"How Much Data is Measured: Sweep Points and Sweep
Count"
*RST:
1001
Example:
LIST:RANG2:POIN 1000
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:RLEVel
<RefLevel>
This command defines the reference level for a spurious emission measurement
range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<RefLevel>
Reference level.
Refer to the data sheet for the reference level range.
*RST:
0 dBm
Default unit: dBm
Example:
LIST:RANG2:RLEV 1DBM
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:SWEep:TIME
<SweepTime>
This command defines the sweep time for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<SweepTime>
Sweep time.
The range depends on the ratios of the span to the RBW and
the RBW to the VBW. Refer to the data sheet for more informa-
tion.
Example:
LIST:RANG2:SWE:TIME 2MS
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:SWEep:TIME:AUTO
<State>
This command turns automatic selection of the sweep time for a spurious emission
measurement range on and off.
Configuring and Performing Measurements