Remote Commands
R&S
®
FPL1000
786
User Manual 1178.3370.02 ─ 11
Parameters:
<Percentage>
Range:
0.001 to 99.999
*RST:
99.999
Default unit: %
Example:
PMET2:DCYC:STAT ON
Activates the duty cycle correction.
PMET2:DCYC:VAL 0.5
Sets the correction value to 0.5%.
Manual operation:
See
[SENSe:]PMETer<p>:FREQuency
<Frequency>
This command defines the frequency of the power sensor.
Suffix:
<p>
.
Power sensor index
Parameters:
<Frequency>
The available value range is specified in the data sheet of the
power sensor in use.
*RST:
50 MHz
Default unit: HZ
Example:
PMET2:FREQ 1GHZ
Sets the frequency of the power sensor to 1 GHz.
Manual operation:
See
[SENSe:]PMETer<p>:FREQuency:LINK
<Coupling>
This command selects the frequency coupling for power sensor measurements.
Suffix:
<p>
.
Power sensor index
Parameters:
<Coupling>
CENTer
Couples the frequency to the center frequency of the analyzer
MARKer1
Couples the frequency to the position of marker 1
OFF
Switches the frequency coupling off
*RST:
CENTer
Example:
PMET2:FREQ:LINK CENT
Couples the frequency to the center frequency of the analyzer
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters