Command Reference
R&S
®
ZNB/ZNBT
1187
User Manual 1173.9163.02 ─ 55
[SENSe<Chn>:]CORRection:SSTate?
Returns the system error correction state label of the active trace in channel
<Chn>
.
The response is a string variable containing the calibration state label in the trace list
('CAL', 'CAI', 'CAL OFF' ..; see
Chapter 5.5.4, "Calibration Labels"
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
See
Usage:
Query only
Manual operation:
See
8.3.14.9
[SENSe:]COUPle...
The
[SENSe:]COUPle...
commands select the sweep mode.
[SENSe<Ch>:]COUPle
<Order>
Determines the order of partial measurements and sweeps.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the sweep mode applies
to all channels in the active recall set.
Parameters:
<Order>
ALL | AUTO | NONE
AUTO - optimized display update: fast sweeps are performed in
alternated mode, slower sweeps in chopped mode
ALL - chopped sweep mode, complete all partial measurements
before proceeding to the next sweep point
NONE - alternated sweep mode on, reverse the order of partial
measurements and sweeps
*RST:
NONE
Example:
COUP NONE
Activate the alternated sweep mode.
TRIG:LINK 'PPO'
Set the triggered measurement sequence equal to one partial
measurement. Each trigger event starts one partial measure-
ment for all sweep points.
Manual operation:
See
8.3.14.10
[SENSe:]DC...
The
[SENSe:]DC...
commands select and configure the DC voltage measurement.
[SENSe<Ch>:]DC<DCInp>:RANGe
<Voltage>
Assigns a DC voltage range to the rear panel input connector DC INPUT <DCInp>.
SCPI Command Reference