Command Reference
R&S
®
ZNB/ZNBT
1181
User Manual 1173.9163.02 ─ 55
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt>
Physical port number, used to select a specific frequency axis in
arbitrary mode (with option R&S
ZNB/ZNBT-K4).
Example:
*RST; :CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:STIM?
Query the stimulus frequencies. The response contains 201 fre-
quency values.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the
active channel.
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:ALL:COUNt?
Returns the total number of port assignments of
all
calibrations.
See
Chapter 5.5.5.4, "Multiple Port Assignments"
on page 171 for background infor-
mation.
Note
:
Use
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?
to get the
number of port assignments of the "current" calibration.
Suffix:
<Ch>
.
Channel number
Example:
See
Chapter 9.2.5.2, "MultiCal (with Calibration Unit)"
Usage:
Query only
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?
Returns the number of port assignments of the "current" calibration, i.e. the calibration
last created using
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
This is particulary useful if the default port assignments were established using
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
.
See
Chapter 5.5.5.4, "Multiple Port Assignments"
on page 171 for background infor-
mation.
Note
:
SCPI Command Reference