GUI Reference
R&S
®
ZNB/ZNBT
295
User Manual 1173.9163.02 ─ 55
Δf is constant over the entire sweep range, if the sweep type is a Lin. Frequency
sweep. For Log. Frequency and Segmented Frequency sweeps, it varies with the
sweep point number m.
Application:
The aperture must be adjusted to the conditions of the measurement. A
small aperture increases the noise in the group delay; a large aperture tends to mini-
mize the effects of noise and phase uncertainty, but at the expense of frequency reso-
lution. Phase distortions (i.e. deviations from linear phase) which are narrower in fre-
quency than the aperture tend to be smeared over and cannot be measured.
Remote command:
CALCulate<Chn>:GDAPerture:SCOunt
Dflt Marker Frmt
Defines the default marker format of the active trace. "Default" means formatted
according to the selected trace format.
New markers are formatted with the trace's "Dflt Marker Frmt"; existing markers are
reformatted if (and only if) their
is set to (Trace) "Default".
For background information on marker formats, see
Remote command:
CALCulate<Chn>:MARKer:DEFault:FORMat
6.4
Scale Softtool
The "Scale" softtool allows you to define how the active trace is displayed in its current
format.
Access:
TRACE – [SCALE] hardkey
6.4.1
Scale Values Tab
Provides the functions for diagram scaling.
Scale Softtool