6.11.1.1
Controls on the Start Cal Tab
●
The "Start Auto Cal" and "Start... (Cal Unit)" buttons are only active if a calibration
unit is connected to the analyzer.
●
If there is more than one channel in the current recall set at the time you initiate a
calibration, the
Channel Selection for Calibration Dialog
is opened.
Start Auto Cal
Performs a fully automatic system error calibration for the active channel, using the
active calibration unit with its factory calibration and auto-detection of ports.
If one port is active, an OSM calibration is performed. If two ports are active, an UOSM
calibration is performed.
Calibration – Start... (Cal Unit)
Depending on the
Multiple Cal in Calibration Wizard
setting, the "Start... (Cal Unit)" but-
ton either opens the
or the
Calibration Unit Wizard (MultiCal)
Calibration – Start... (Manual)
Depending on the
Multiple Cal in Calibration Wizard
setting, the "Start... (Cal Unit)" but-
ton either opens the
or the
Calibration Setting Wizard (Multi-
Calibration – Repeat...
Re-enters the data acquisition step of the current calibration, restoring the related
channel settings. This provides a convenient way to repeat or correct an existing cali-
bration without necessarily repeating all measurements.
Scalar Power Cal – Power Cal...
Cal Softtool