GUI Reference
R&S
®
ZNB/ZNBT
570
User Manual 1173.9163.02 ─ 55
Access:
> "Offset"
Figure 6-47: Offset Embed dock widget: Offset panel
The "Offset" panel can also be activated by selecting the
or
. Refer to its description for additional functions and remote commands.
Active
The checkbox in the "Active" column activates/deactivates the configured length and
loss parameters for the respective
(i.e. adds/removes them to/from the calculation
flow) without changing the parameter values.
Remote command:
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe]
Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd Freq / Loss at 2nd
Freq / 2nd Freq for Loss
Defines the one-way loss parameters of the transmission line at the selected port.
If the loss at DC and one additional frequency
f
1
is known, the loss at frequency f is
approximated by:
If, in addition, the loss at a second frequency
f
2
is known and "Use 2nd Freq" is
enabled, then the loss is approximated by:
See also
Chapter 5.6.1.2, "Definition of Loss Parameters "
Note:
The entered parameters define the loss for a signal traveling in
one direction
through the transmission line. To account for the propagation in both directions, the
magnitude shift of a reflection parameter due to a given loss is twice the magnitude
shift of a transmission parameter. See also
Chapter 5.6.1.6, "Application and Effect of
Remote command:
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet
[SENSe<Ch>:]CORRection:LOSS<PhyPt>
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:FREQuency
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:STATe
Offset Embed Softtool