Command Reference
R&S
®
ZNB/ZNBT
1218
User Manual 1173.9163.02 ─ 55
[SENSe<Ch>:]ROSCillator[:SOURce]
<Source>
Selects the source of the reference oscillator signal.
Parameters:
<Source>
INTernal | EXTernal
INTernal – internal 10 MHz reference oscillator
EXTernal – external reference clock
The frequency of the external reference clock can be defined
using
[SENSe:]ROSCillator:EXTernal:FREQuency
*RST:
INTernal
Example:
See
[SENSe:]ROSCillator:EXTernal:FREQuency
Manual operation:
See
[SENSe:]ROSCillator:EXTernal:FREQuency
<ExtClockFreq>
Specifies or queries the frequency of the external reference oscillator.
Parameters:
<ExtClockFreq>
Frequency of the external reference clock signal.
Range:
See the data sheet of your analyzer.
*RST:
10 MHz
Default unit: Hz
Example:
ROSC EXT
Select the external reference clock as clock source.
ROSC:EXT:FREQ 10MHz
Specify the frequency of the external reference clock.
ROSC:EXT:FREQ?
Query the frequency of the external reference oscillator. The
response is 10000000 Hz.
Manual operation:
See
8.3.14.18
[SENSe:]SEGMent<Seg>...
The
[SENSe:]SEGMent<Seg>...
commands define all channel settings for a seg-
mented frequency sweep. A segmented sweep is activated via
SEGMent
.
The commands in this subsystem do not accept the step parameters UP and DOWN.
Numeric values can be entered directly or using the DEFault, MINimum, MAXimum
parameters.
..................................................................................1219
[SENSe<Ch>:]SEGMent<Seg>[:STATe]
......................................................................... 1220
..............................................................................1220
SCPI Command Reference