Use the REF OUT connector at the rear of the instrument to synchronize other
devices (e.g. signal generators or a second R&S
ZNB/ZNBT network analyzer).
●
External:
The analyzer synchronizes to an external reference clock via the REF IN
connector at the rear of the instrument.
The external reference signal has to meet the specifications of the data sheet; its
frequency must be specified in the "Ext Frequency" field. The internal reference
signal is synchronized to the external signal. The external signal is also looped to
REF OUT, so that it can be reused to synchronize other devices.
Remote command:
[SENSe<Ch>:]ROSCillator[:SOURce]
Ext Frequency
Specifies the frequency of the external reference clock signal at REF IN.
Remote command:
[SENSe:]ROSCillator:EXTernal:FREQuency
6.17.3
Remote Settings Tab
Configures the remote control interfaces of the R&S
ZNB/ZNBT.
6.17.3.1
Controls on the Remote Settings Tab
IP Address
Displays the current IP4 address of the R&S
ZNB/ZNBT. By default, the analyzer is
configured to use dynamic TCP/IP configuration (DHCP) and obtain all IP address
information automatically. See
Chapter 4.1.12.1, "Assigning an IP Address"
The button opens the Windows "Network Connections" system config dialog that
allows you to configure the IP4 (and IP6) settings of your instrument.
GPIB Address
Defines the analyzer's GPIB address. The address must be in the range between 0
and 30.
Setup Softtool