Command Reference
R&S
®
ZNB/ZNBT
1183
User Manual 1173.9163.02 ─ 55
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
<TestPort1>, <TestPort2>...
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:
, but initially defines the default port assignments only by their
underlying test port set(s); the connected calibration unit ports are
auto-detected
at
the start of the corresponding calibration sweep.
The number of created assignments can be queried using the
]CORRection:COLLect:AUTO:ASSignment:COUNt?
command.
See
Chapter 5.5.5.4, "Multiple Port Assignments"
on page 171 for background infor-
mation.
Suffix:
<Ch>
.
Setting parameters:
<TestPort1>
<TestPort2>
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DELete:ALL
Deletes all port assignments (see
Chapter 5.5.5.4, "Multiple Port Assignments"
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Event
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
Starts the automatic calibration sweep for the indicated channel and port assignment.
A complete, valid set of port assignments must be defined before you can initiate a cal-
ibration. See
Chapter 5.5.5.4, "Multiple Port Assignments"
on page 171 for back-
ground information.
Suffix:
<Ch>
.
Number of the calibrated channel
<Asg>
Number of the port assignment
SCPI Command Reference