Command Reference
R&S
®
ZNB/ZNBT
841
User Manual 1173.9163.02 ─ 55
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<Length>
Length in inches
Options:
R&S
ZNB/ZNBT-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:M1L:MEASure
CALCulate:FMODel:DELT<Ph_pt>:M2L:MEASure
CALCulate:FMODel:DELT<Ph_pt>:M3L:MEASure
Measures the full set of S-parameters at the active ports (see
Del:DELT<Ph_pt>:M1L|M2L|M3L[:STATe]
).
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage:
Event
Options:
R&S
ZNB/ZNBT-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:M1L[:STATe]
<Boolean>
CALCulate:FMODel:DELT<Ph_pt>:M2L[:STATe]
<Boolean>
CALCulate:FMODel:DELT<Ph_pt>:M3L[:STATe]
<Boolean>
Defines the active ports for the 1st/2nd/3rd line Delta-L measurement (
late:FMODel:DELT<Ph_pt>:M1L|M2L|M3L:MEASure
).
For each measurement, either two or four ports must be active. If the 1-length method
is used (
CALCulate:FMODel:DELT<Ph_pt>:METHod M1L
active for the first line.
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean>
ON (1)
Port <Ph_pt> is active in the respective measurement
OFF (0)
Port <Ph_pt> is inactive in the respective measurement
Options:
R&S
ZNB/ZNBT-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:MEASurement
<Measurement>
Defines whether single-ended or balanced analysis is used in Delta-L characteriza-
tions.
SCPI Command Reference