background image

OASIS   

 

H.Z.02290025 

A product recommended by State Administration of Grain for popularization 
One of the first batches of products having passed a nationwide survey by State 
Administration of Grain 
ISO9001: 2000 certified 

 

 

LDS-IG 

Computerized Moisture Meter 

User’s Manual 

 

 

 

Quick, accurate, convenient 
AC/DC power supply compatible 
High-brightness backlight LCD, clear and energy-saving 
Multipoint calibration, error correction 
Low power consumption, automatic shutdown 
Automatic weighing and temperature compensation 
Volume-weight conversion display 

 
 
 

 

Shanghai QingPu Oasis Inspection Technologies Co., Ltd

Summary of Contents for LDS-IG

Page 1: ... ISO9001 2000 certified LDS IG Computerized Moisture Meter User s Manual Quick accurate convenient AC DC power supply compatible High brightness backlight LCD clear and energy saving Multipoint calibration error correction Low power consumption automatic shutdown Automatic weighing and temperature compensation Volume weight conversion display Shanghai QingPu Oasis Inspection Technologies Co Ltd ...

Page 2: ... Lower edge of blanking cylinder the place for sample to be placed during measurement Funnel Sample container Switch for discharge gate Discharge gate Schematic drawing of LCD panel display when the dry battery is under voltage Indicate that the value is the average of the measure Indicate different working statuses of the moisture meter Indicate the corresponding unit of measurement result Indica...

Page 3: ...ry code list attached to this User s Manual and chooses the corresponding category code for moisture measurement Note when the meter is used for trade settlement and safe storage which needs higher precision we strongly recommend that you adopt standard sample of the category to be measured and conduct calibration and error correction of the meter so as to ensure measuring precision and safeguard ...

Page 4: ...lue from the standard value For instance the moisture value measured by the meter is 13 6 but actual moisture value to be displayed is 14 0 So correction value is 0 4 which means that the calibration is 0 4 upward If the correction value is negative that means the downward calibration is necessary 2 Enter into correction status take the sample out of the meter press and hold the Category key and r...

Page 5: ... code of the category to be calibrated the previous calibration parameter under this code will be overlaid 4 Enter into calibration status press and hold the 确定 OK key approx 5 6s and release it after hearing the buzz The character Calibration flickers at the bottom left corner of the LCD panel and is displayed which means that the meter has entered into the calibration status and the first standa...

Page 6: ...en press and hold the Category key until the buzz is heard The meter displays flickering which means the default parameter has been restored Shut down the meter and exit the restoration status VII Status Display The meter has self check functionality upon startup which can give prompt signals according to different working status shows that there is a sample in the sensor during startup or somethi...

Page 7: ...g put into the meter for measurement After showing the moisture value press the 确定 OK key to display the sample weight Then press the 确定 OK key again to starts automatic conversion shortly after which the volume weight of the sample will be displayed e g This value of course is for reference only 5 Communication interface used for special occasions Please contact us for its technical parameters an...

Page 8: ...racy due to unexpected factors which can be calibrated in the following manner a Horizontally place the meter in shutdown state and take everything out of the sensor Press and hold the Category key then press the Power key for startup and the meter will buzz Release the Category key and the meter will display a figure indicating the meter is under balance calibration status b Press the 确定 OK key a...