Command Reference
R&S
®
ZNB/ZNBT
1346
User Manual 1173.9163.02 ─ 55
Usage:
Setting only
8.3.19
TRIGger Commands
8.3.19.1
TRIGger[:SEQuence]...
The
TRIGger[:SEQuence]...
commands synchronize the analyzer's measurement
sequences to external events. These events are indicated to the VNA by sending trig-
ger signals via EXT TRIG IN or the USER PORT.
TRIGger<Ch>[:SEQuence]:HOLDoff
.............................................................................. 1346
................................................................................... 1346
TRIGger<Ch>[:SEQuence]:MULTiple:HOLDoff
................................................................1347
TRIGger<Ch>[:SEQuence]:MULTiple:SLOPe<Num>
....................................................... 1347
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce
............................................................... 1348
................................................................................ 1349
..............................................................................1349
TRIGger<Ch>[:SEQuence]:HOLDoff
<DelayTime>
Sets a delay time between the trigger event and the start of the measurement ("Trigger
Delay").
Suffix:
<Ch>
.
Channel number.
Parameters:
<DelayTime>
Delay time.
Range:
0 s to 13680 s
Increment: 10 ms
*RST:
0 s
Default unit: s
Example:
TRIG:SOUR MAN
Activate external trigger source.
TRIG:HOLD UP
Set a delay time of 10 ms.
Manual operation:
See
TRIGger<Ch>[:SEQuence]:LINK
<MeasSequence>
Selects the triggered measurement sequence. The identifier for the sequence is a
string variable.
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasSequence>
Triggered measurement sequence, string variable.