Command Reference
R&S
®
ZNB/ZNBT
833
User Manual 1173.9163.02 ─ 55
CALCulate:FMODel:SFD<Ph_pt>:COUPon:MEASure
For a fixture modeling with the SFD tool, this command starts the measurement of the
coupon at the active ports (see
CALCulate:FMODel:SFD<Ph_pt>:COUPon[:
).
The coupon type can be selected using
.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage:
Event
Manual operation:
See
"Measure / Measure Open / Measure Short"
CALCulate:FMODel:SFD<Ph_pt>:COUPon:MEASure:FILename
<String>
For a fixture modeling with the SFD tool, this command loads the coupon properties
from a Touchstone file.
The coupon type can be selected using
.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<String>
Path to the Touchstone file, either absolute or relative to the cur-
rent directory (see
Manual operation:
See
"Load File / 1x Open Preset / 1x Short Preset "
CALCulate:FMODel:SFD<Ph_pt>:COUPon:TYPE
<CouponType>
Sets/gets the coupon type to be measured for a fixture modelling with the SFD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<CouponType>
SYMMetric2x | OPEN1x | SHORt1x
Manual operation:
See
CALCulate:FMODel:SFD<Ph_pt>:COUPon[:STATe]
<Boolean>
For a fixture modeling with the SFD tool, this command allows to specify the ports to
which the test coupon is connected.
This has to be defined before measuring the test coupon (using
).
SCPI Command Reference