Sweep Active Channel
Apply the sweep control settings to the active channel only. The number of sweeps in a
"Single" sweep sequence is equal to the number of "Sweeps" in the active channel.
The table lists all channels in the active recall set and allows you to define individual
numbers of sweeps for all channels. When a new channel is selected, the analyzer
uses its specific number of sweeps.
Remote command:
INITiate<Ch>[:IMMediate]:SCOPe
SINGle
6.11
Cal Softtool
The "Cal" softtool provides all functions related to system error calibration, scalar
power calibration, and "SMARTerCal".
Access:
CHANNEL – [CAL]
6.11.1
Start Cal Tab
The "Start Cal" tab provides access to all functions for automatic or manual calibration.
Calibration of the R&S
ZNB/ZNBT is a fully guided process.
Background information
Refer to the following sections:
●
●
Chapter 5.5.5, "Automatic Calibration"
●
Chapter 5.5.6, "Scalar Power Calibration"
●
●
It is possible to set up, perform and activate multiple system error corrections or
SMARTerCals in a single calibration procedure. However, this
feature is disabled by default and has to be activated in the "Calibration"
tab of the "System Config" dialog.
Cal Softtool