Command Reference
R&S
®
ZNB/ZNBT
1132
User Manual 1173.9163.02 ─ 55
[SENSe<Ch>:]CORRection:METas:CABLe:REConnection
[<Ports>, <Ports>...]
While a METAS calibration is active for the related channel (see
on page 1133), this command notifies the METAS
VNA Tools II about additional connector reconnections at the specified ports.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Ports>
<integer>
First port number, where connectors were reconnected
ALL
All ports were reconnected
NONE
Clears the connector reconnection history in the METAS project
<Ports>
Additional port numbers, where connectors were reconnected
Usage:
Setting only
Options:
R&S
ZNB-K50
Manual operation:
See
"Add Conn. Reconnection/Add Cable Movement"
[SENSe<Ch>:]CORRection:METas:CABLe:TYPE
<CableNameFile>[, <Ports>,
<Ports>...]
Selects the METAS cable type that is used at the respective ports.
Use
[SENSe:]CORRection:METas:CABLe:CATalog?
to list the available cable
types.
Suffix:
<Ch>
.
Channel number
The related channel must be prepared for a METAS calibration
using
[SENSe<Ch>:]CORRection:COLLect:METHod:
<CalName>, MTOSm | MUOSm | MOSM |
MTRL, ...
.
Setting parameters:
<CableNameFile>
Name of a METAS cable model file (without
*.cable
exten-
sion) in the database of the local METAS VNA Tools II installa-
tion.
<Ports>
First port number or
ALL
, if the same cable type is used for all
ports.
The latter is equivalent to omitting all <Ports> parameters.
<Ports>
Additional port numbers
Usage:
Setting only
Options:
R&S
ZNB-K50
SCPI Command Reference