background image

Keysight 

Add 4-Port Capability 
Upgrade Kit

Installation 

Note

To Upgrade PNA N5221B or N5222B Option 21

7

 to Option 41

7

Upgrade Kit Order Number: N5221BU- 6

17

 and N5222BU- 61

7

Keysight Kit Number: N5222-6011

8

Summary of Contents for N5222-60118

Page 1: ...eysight Add 4 Port Capability Upgrade Kit Installation Note To Upgrade PNA N5221B or N5222B Option 217 to Option 417 Upgrade Kit Order Number N5221BU 617 and N5222BU 617 Keysight Kit Number N5222 6011...

Page 2: ...S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license set...

Page 3: ...on 217 2 port analyzer to a N5221B or N5222B Option 417 4 port analyzer by adding an additional source an additional source synthesizer an additional mixer brick two additional receiver couplers two a...

Page 4: ...Web at http www keysight com find assist If you do not have access to the Internet please contact your Keysight field engineer If You Have Problems With the Upgrade Kit Contents Keysight stands behin...

Page 5: ...tware Manager http www keysight com find softwaremanager To complete the request you will need to gather the following information From the certificate Order number The PNA contains extremely sensitiv...

Page 6: ...umber use the following steps 1 Go to www keysight com 2 In the Search box enter the model number of the analyzer Ex N5242A and click Search 3 Click Technical Support Manuals 4 Click Service Manual 5...

Page 7: ...pen end torque wrench set to 6 in lbs 0 68 N m 1 5 8 in 16 mm nutsetter or open end torque wrench set to 21 in lbs 2 38 N m 1 1 in 25 4 mm torque wrench set to 72 in lbs 8 15 N m 1 Use a 5 16 in torqu...

Page 8: ...st port 3 source attenuator 2 33321 60077 A36 Test port 4 source attenuator A43 Test port 3 receiver attenuator 2 33321 60078 A44 Test port 4 receiver attenuator Dress panel overlay 4 Port 1 N5240 000...

Page 9: ...ard P5 to W6 on bottom of analyzer 1 N5222 20062 W6 RF cable W5 to A26 port 3 receiver coupler 1 N5222 20041 W7 RF cable A8 source 2 board P3 to W8 on bottom of analyzer 1 N5222 20063 W8 RF cable W7 t...

Page 10: ...A12 SPAM board J5 1 N5242 60015 W67 RF cable A10 frequency reference board J7 to bottom of A13 13 5 GHz source 2 synthesizer board J5 4 port only 1 N5242 60030 W75 RF cable A26 port 3 receiver couple...

Page 11: ...urce attenuator 1 N5242 60008 Extra quantities of items such as protective plastic caps screws cable ties and cable clamps may be included in this upgrade kit It is normal for some of these items to r...

Page 12: ...A27 Receiver Coupler Assemblies Step 12 Assemble the A35 and A36 Source Attenuators and the A43 and A44 Receiver Attenuators Step 13 Install the A35 and A36 Source Attenuators and the A43 and A44 Rec...

Page 13: ...ctions on the Software Entitlement Certificate supplied to obtain a license key file for installation of this upgrade Refer to License Key Redemption on page 5 Verify that the model number serial numb...

Page 14: ...e an image showing the location of these cables click the Chapter 6 bookmark Bottom RF Cables Standard 2 Port Configuration Option 200 in the PDF Service Guide2 These cables may be discarded they will...

Page 15: ...r attenuator to A23 mixer brick A W13 N5222 20005 A25 port 1 receiver coupler to A33 reference mixer switch W35 N5222 20033 A33 reference mixer switch to front panel REF 1 SOURCE OUT W36 N5222 20032 R...

Page 16: ...16 Installation Note N5222 90118 Installation Procedure for the Upgrade Figure 1 A24 Mixer Brick Assembly...

Page 17: ...Installation Note N5222 90118 17 Installation Procedure for the Upgrade Figure 2 A24 Mixer Brick Block Assembly cont...

Page 18: ...tion Procedure for the Upgrade Step 9 Assemble the A26 and A27 Receiver Coupler Assemblies Follow the instructions shown in Figure 3 New parts are listed in Table 1 on page 8 of this document Figure 3...

Page 19: ...llow the instructions shown in Figure 4 New parts are listed in Table 1 on page 8 of this document Figure 4 Mixer Brick Block and Receiver Coupler Assemblies Installation Step 11 Install the A26 and A...

Page 20: ...cedure for the Upgrade Step 12 Assemble the A35 and A36 Source Attenuators and the A43 and A44 Receiver Attenuators Follow the five instructions shown in Figure 5 Figure 5 A35 A36 Source Attenuators A...

Page 21: ...ocedure for the Upgrade Step 13 Install the A35 and A36 Source Attenuators and the A43 and A44 Receiver Attenuators Follow the instructions shown in Figure 6 Figure 6 A35 A36 Source Attenuators Assemb...

Page 22: ...rom the PNA For instructions click the Chapter 7 bookmark Removing and Replacing the A29 A32 Test Port Couplers in the PDF Service Guide2 2 Using pliers remove the adhesive bumper on the A29 test port...

Page 23: ...from the 2 port test set front plate of the PNA Keep the screws and LED boards for reuse later 2 Remove the 12 bulkhead connectors nuts and washers from the 2 port front plate to reuse later 3 Remove...

Page 24: ...new bulkhead connectors included in the kit for the remainder of this step 2 From the back side of the test set front plate insert a bulkhead connector into a hole in the plate 3 Install 1x washer an...

Page 25: ...ote N5222 90118 25 Installation Procedure for the Upgrade Step 17 Install the 4 Port Coupler Plate Assembly to the Deck Follow the four instructions shown in Figure 10 Figure 10 Coupler Plate Assembly...

Page 26: ...6 Installation Note N5222 90118 Installation Procedure for the Upgrade Step 18 Assemble the A8 26 5 GHz Source 2 Assembly Follow the three instructions shown in Figure 11 Figure 11 A8 Source 2 Assembl...

Page 27: ...ation Note N5222 90118 27 Installation Procedure for the Upgrade Step 19 Install the A8 26 5 GHz Source 2 Assembly Follow the three instructions shown in Figure 12 Figure 12 A8 Source 2 Assembly Insta...

Page 28: ...ly connected to J5 of the A13 source 2 synthesizer board Step 21 Install the Test Set Cables Flexible Cables Required for Upgrading to an Option 401 PNA Install the following gray flexible cables in t...

Page 29: ...4 coupler to front panel Port 4 CPLR ARM W80 N5222 20020 A36 port 4 source attenuator to front panel Port 4 SOURCE OUT W50 N5222 20027 REF 4 RCVR R4 IN to A24 mixer brick R4 W92 N5222 20026 A44 port 4...

Page 30: ...20033 W17 N5222 20023 A26 port 3 receiver coupler to front panel REF 3 SOURCE OUT W75 N5222 20002 A26 port 3 receiver coupler to A35 port 3 source attenuator W90 N5222 20006 A43 port 3 receiver attenu...

Page 31: ...ure for the Upgrade Figure 14 Location of Cable Clamps to Secure W49 W48 N5222 20101 A33 reference mixer switch to A23 mixer brick R1 As shown in Figure 15 on page 32 install three cable ties part num...

Page 32: ...reference mixer switch If the screws that attach the reference mixer switch to the test set deck were loosened torque these screws now to 9 in lbs W10 reuse N5222 20043 W9 to A28 port 2 receiver coupl...

Page 33: ...eiver attenuator 8121 0982 A19 test set motherboard J207 to A44 port 4 receiver attenuator N5242 60008 A19 test set motherboard J202 to A35 port 3 source attenuator N5242 60008 A19 test set motherboar...

Page 34: ...ont Panel Assembly For instructions on reinstalling the front panel assembly click the Chapter 7 bookmark Removing and Replacing the Front Panel Assembly in the PDF Service Guide1 Be sure to install t...

Page 35: ...e W30 front panel jumper cables N5222 20091 use 6 new jumpers and reuse 6 old jumpers To see an image of the front panel jumper cables click the Chapter 7 bookmark Removing and Replacing the Front Pan...

Page 36: ...cedure 1 To start the Keysight License Manager press Start Keysight License Manager Keysight License Manager A Keysight License Manager dialog box will appear 2 Right click the on the desired option a...

Page 37: ...m Keysight which contain license file attachments These emails are a result of Step 1 Obtain a Keyword and Verify the Information on page 13 2 Copy the license file s from the email s to the root dire...

Page 38: ...Network Analyzer program is again running 1 On the analyzer s Help menu press About Network Analyzer 2 Verify that 417 are listed but 217 is not listed after Options in the display Click OK Attempting...

Page 39: ...kmark Tests and Adjustments in the PDF Service Guide1 After the specified adjustments have been performed the analyzer should operate and phase lock over its entire frequency range Operator s Check Pe...

Page 40: ...e PNA for the User 1 If necessary reinstall front jumper cables 2 Install the cable guards pushing them over the front jumper cables until the cushioning material touches the front panel of the PNA 3...

Page 41: ...Installation Note Xxxxx xxxxx 3...

Page 42: ...This information is subject to change without notice Keysight Technologies 2000 2017 Edition 2 May 2017 www keysight com...