background image

   

L1 

Lactate Monitoring System 

 
 

Owner’s Manual 

 

 

Owner’s Manual 

 
 

Summary of Contents for L1

Page 1: ...L1 Lactate Monitoring System Owner s Manual Owner s Manual...

Page 2: ...ustomer service or place of purchase Intended Use This system is intended for use outside the body in vitro diagnostic use by people with suspecting of having lactic acidosis at home and by healthcare...

Page 3: ...Keep the device and testing equipment away from young children Small items such as the battery cover batteries test strips lancets and vial caps are choking hazards 7 Use of this instrument in a dry e...

Page 4: ...Your Meter 7 Checking the Code Number 7 QUALITY CONTROL TESTING 8 When Should the Control Solution Test be Performed 8 Performing a Control Solution Test 8 TESTING WITH BLOOD SAMPLE 10 Preparing the L...

Page 5: ...aring for Your Test Strips 16 Cleaning Your Lancing Device 17 Important Control Solution Information 18 SYSTEM TROUBLESHOOTING 19 Result Readings 19 Error Messages 19 Troubleshooting 21 DETAILED INFOR...

Page 6: ......

Page 7: ......

Page 8: ...or continue to get results which are higher or lower than usual follow the treatment advice of your healthcare professional Use only fresh whole blood samples to test your lactate Using other substan...

Page 9: ...dicator Test Strip Ejector Eject the used strip by pushing up this button Reset Button Data Port M Button M Enter the meter memory Test Strip Slot Insert test strip here to turn the meter on for testi...

Page 10: ...ess M to turn on the meter Press or to select from the modes below Indicator Meaning You can do this Lactate Measuring Mode Perform your lactate tests Memory Recall Mode Review your lactate test resul...

Page 11: ...G THE METER Before using your meter for the first time or if you change the meter battery you should check and update these settings Entering the Setting Mode a Start with the meter off no test strip...

Page 12: ...displays the complete time and Time is set message 5 Deleting the memory To delete all the results press or to switch to yes and press M The meter displays Memory cleared which indicates that all dat...

Page 13: ...de after a test strip is inserted press or to select a suitable measuring mode when the meter displays the list Press M and proceed with a lactate testing BEFORE TESTING Calibration You must calibrate...

Page 14: ...ial or box of foil package before you proceed b If it matches you can proceed with your test If the codes do not match please stop testing and contact Customer Service for help NOTICE The code number...

Page 15: ...ay not be included in the kit please check the contents on your product box They can be purchased separately Please make sure you have those items needed for a lactate test beforehand Performing a Con...

Page 16: ...nt hole of the test strip to touch the drop Once the confirmation window fills completely the meter will begin counting down NOTE To avoid contaminating the control solution do not directly apply cont...

Page 17: ...e lactate value obtained Blood from a site that has not been rubbed exhibits a measurably different lactate concentration than blood from the finger When the puncture site was rubbed prior to blood ex...

Page 18: ...ply the sample k Gently apply the drop of blood to the absorbent hole of the test strip at a tilted angle Confirmation window should be completely filled if enough blood sample has been applied Do NOT...

Page 19: ...ically 6 Eject the used test strip l Eject the test strip by pushing the eject button on the top The meter will switch itself off automatically Always follow the instructions in the lancing device ins...

Page 20: ...pressing or Press M to enter The first reading you see is the last lactate result along with test number date time and the measuring mode 2 Press or to recall the test results stored in the meter eac...

Page 21: ...Health Care Software System please visit the TaiDoc s website www taidoc com 2 Connecting to a personal computer Connect the cable to a USB port on your computer With the meter switched off connect th...

Page 22: ...rsonal computer 3 The recharge symbol shows when battery is recharging After the battery is fully charged remove the USB cable and then the meter will automatically turn off NOTE Recharging the batter...

Page 23: ...humidity Meter Disposal The used meter should be treated as contaminated that may carry a risk of infection during measurement The batteries in this used meter should be removed and the meter should...

Page 24: ...choking hazard If swallowed promptly see a doctor for help For further information please refer to the test strip package insert Cleaning Your Lancing Device Cleaning Procedures 1 Take out one alcoho...

Page 25: ...months It is recommended that the control solution test be done at room temperature 20 C to 25 C 68 F to 77 F Make sure your control solution meter and test strips are at this specified temperature r...

Page 26: ...ediately Error 23 Appears when there is a system error Repeat the test with a new test strip If the meter still does not work please contact the customer service for assistance Error 24 Appears when a...

Page 27: ...please the customer service for assistance Error 45 Appears when there is insufficient blood volume Review the instructions and repeat test with a new strip If the problem persists please contact the...

Page 28: ...ion Repeat the test with a new test strip Apply sample only when flashing appears on the display Defective meter Please contact customer services 3 If the control solution testing result is out of ran...

Page 29: ...ers Co Please consult your doctor to determine a target range that works best for you SYMBOL INFORMATION SYMBOL REFERENT SYMBOL REFERENT In vitro diagnostic medical device Caution consult accompanying...

Page 30: ...off after 3 minutes without action Temperature Warning Operating Condition 10 C to 40 C 50 F to 104 F below 85 R H non condensing Meter Storage Transportation Conditions 20 C to 60 C 4 F to 140 F belo...

Page 31: ...For self testing TaiDoc Technology Corporation B1 7F No 127 Wugong 2nd Rd Wugu Dist 24888 New Taipei City Taiwan www taidoc com MedNet GmbH Borkstra e 10 48163 M nster Germany...

Reviews: