The generated error code can be retrieved via remote control command
SOURce<Ch>:GROup:SIMultaneous:FOFFset:CONDition?
. The configuration
problems listed above are indicated as error codes -8 "invalid offset or frequency spac-
ing" and -6 "no simultaneous mode possible".
Remote command:
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:CONDition
6.12.3
Mode Tab
Optimizes the measurement process.
6.12.3.1
Controls on the Mode Tab
The controls on the "Mode" tab are organized in sections "Driving Mode", "Image
Suppr." and "AGC Mode".
Driving Mode
Determines the order of partial measurements and sweeps.
●
In "Auto" mode, the analyzer optimizes the display update: Fast sweeps are per-
formed in "Alternated" mode, slower sweeps in "Chopped" mode.
●
In "Alternated" mode, a partial measurement is performed at all sweep points (par-
tial sweep) before the hardware settings are changed. The next partial measure-
ment is carried out in an additional sweep.
This mode is usually faster than "Chopped" mode.
●
In "Chopped" mode, the analyzer completes the necessary sequence of partial
measurements at each sweep point and obtains the result (measurement point)
before proceeding to the next sweep point. A trace is obtained from the beginning
of the sweep.
Channel Config Softtool