47
4 Mounting
4.4.1
Rotation of the transmitter housing
1.
Observe chapter 4.4 "Notes on opening and closing the housing", page 46!
2. Loosen the Allen screws (3) on the front and rear, but do not remove them completely.
3. Loosen the screws (2) and rotate the transmitter housing to the left or right by 90°.
4. Before re-starting, tighten all screws (2) and Allen screws (3) on the transmitter housing.
4.4.2
Turning the LCD display
Depending on the installation position, the LCD display or transmitter housing can be rotated to allow for
horizontal read-off.
1. Switch off voltage supply.
2. Remove case lid
see chapter 11.4.2 "Replacing the transmitter", page 176.
3. Slightly pull back the rotation lock (3) and rotate the transmitter plug-in module with LCD display (1)
to the left or right by 90° until the rotation lock (3) locks back into place.
4. Relocate the case lid and screw down tightly before re-starting.
Dual-compartment housing
Single-compartment housing
(1)
Transmitter housing
(2)
Screw
(3)
Allen screw
(1)
Transmitter with LCD display
(2)
Housing for transmitter plug-in module
(3)
Rotation lock
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......