139
8 Parameterization
Italics = Parameter can only be viewed in the "Advanced" password level
Gas Bubble Detector
The gas bubble detector can be used in the nominal width range DN 10 to 300. More information
see
chapter 9 "Advanced diagnostics functions", page 153.
Adjust Gas Bubble
Prior to starting the (manual/automatic) calibration, make sure that:
•
There must be no flow through the sensor (close all valves, shut-off devices, etc.).
•
The sensor must be completely filled with the medium to be measured.
Italics = Parameter can only be viewed in the "Advanced" password level
Coating Detector
The electrode deposit detector can be used in the nominal width range DN 10 to 300. More information
see chapter 9 "Advanced diagnostics functions", page 153.
Menu/parameter
Value range
Description
Diagnostics/Diagnostics Control/
Gas Bubble Detector
a
Gas Bubble Detector
ON/OFF
Activates the "Gas Bubble Detector" function
Default setting: OFF
Adjust Gas Bubble
The Gas Bubble Detector must be calibrated to the
local conditions. Start automatic gas bubble detection
calibration.
Threshold Gas Bubble
Setting the switching threshold. If the current value is
above the set switching threshold, a message is out-
put on the LCD display and the alarm is activated at
the digital output, if configured accordingly.
Gas Bubble Value
Display only
Display the current Gas Bubble Value
a
Only available for JUMO flowTRANS MAG S02/H02
?
?
Menu/parameter
Value range
Description
Diagnostics/Diagnostics Control/
Coating Detector
a
Coating Detector
ON/OFF
Activates the "Coating Detector" function
Default setting: OFF
One Shot Coating
The electrode deposit is measured cyclically at spec-
ified intervals. A one-off measurement can be initiat-
ed here.
Coating Value QE1
Display only
Current deposit value for electrode 1
Coating Value QE2
Display only
Current deposit value for electrode 2
Coating QE Min. Alarm
0 to 100,000
Setting of the minimum limit value for detection of de-
posits. The smaller value of QE1, QE2 triggers an
alarm if below the threshold.
Default setting: 0
Coating QE Max. Alarm
0 to 100,000
Setting of the minimum limit value for detection of de-
posits. The smaller value of QE1, QE2 triggers an
alarm if below the threshold.
Default setting: 100.000
a
Only available for JUMO flowTRANS MAG S02/H02
?
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......