89
7 Startup
31. Use
to call up the edit mode.
32. Use
or
to select the desired alarm mode.
33. Use
to confirm your selection.
34. Use
to call up the edit mode.
35. Use
or
to set the desired current for low alarm.
36. Use
to confirm your setting.
37. Use
to call up the edit mode.
38. Use
or
to set the desired current for high alarm.
39. Use
to confirm your setting.
40. Use
to start the automatic zero point adjustment for the system.
NOTE!
Prior to starting the zero point adjustment, make sure that:
There must be no flow through the sensor (close all valves, shut-off devices, etc.).
The sensor must be completely filled with the medium to be measured.
41. Use
to call up the edit mode.
42. Use
or
to enter the signal cable length.
Easy Setup
Low Alarm
V
alue
3.5000 mA
Next
Edit
Easy Setup
High Alarm
V
alue
2
1.800 mA
Next
Edit
Easy Setup
System Zero
█ █ █ █ █ █ █
Next
OK
Easy Setup
Cable length
5.0000 m
Next
Edit