10 Error messages
166
S110.035
Operation
Sensor setup Cal-Status
Sensor is uncalibrated or Cal-Sta-
tus is not set to "calibrated".
Contact manufacturer
Set Cal-Status to calibrated
S108.044
Operation
Pulse output is cutted off
Incorrect configuration
In the "Easy Setup" menu, reduce
the "Pulses per unit" value.
Check pulse out configura-
tion
Error no./
area
Text in the LCD display
Cause
Remedy