9 Advanced diagnostics functions
9
Advanced diagnostics functions
9.1
General Information
9.1.1
Partial filling detection
Optionally, a measuring electrode (TFE electrode) is available for detecting partial filling of the sensor.
The partial filling alarm is output via the programmable digital output.
Conditions for using the function:
•
Nominal width as of DN 50 (2 inch) with the Design Level "B" sensor
•
Maximum signal cable length in case of remote mount design type: 200 m (656 ft).
•
For this function, the conductivity of the measurement medium must be between 20 µS/
cm and 20,000 µS/cm.
•
The function is only available for JUMO flowTRANS S/H devices without Ex-protection or with Ex-
protection for zone 2.
Additional installation conditions:
•
The sensor must be mounted horizontally with the connection box facing up.
9.1.2
Gas bubble detection
Gas bubbles in the measurement medium are detected by reference to a configurable maximum thresh-
old. When the threshold is exceeded, an alarm is triggered via the programmable output, depending on
the configuration.
Conditions for using the function:
•
JUMO flowTRANS MAG S02:
The function is available in the nominal width range DN 10 to 300 (3/8 to 12 inch).
•
JUMO flowTRANS MAG H02:
The function is available in the nominal width range DN 10 to 100 (3/8 to 4 inch).
•
Maximum signal cable length in case of remote mount design type: 50 m (164 ft).
•
For this function, the conductivity of the measurement medium must be between 20 µS/
cm and 20,000 µS/cm.
Additional installation conditions:
•
The sensor can be mounted horizontally or vertically. Vertical mounting is preferable.
IMPORTANT (NOTE)!
•
The advance diagnostics functions are only available with the JUMO flowTRANS S02/H02 devices.
•
The "Partial filling detection" function is not available with the JUMO flowTRANS H02.
•
When using the advanced diagnostics functions, note that a pre-amplifier must not exist on the ex-
ternal sensor.
•
To facilitate initial startup, the advance diagnostics functions are disabled as the default setting.
•
To use the advance diagnostics functions, a "Startup fingerprint" must be created on startup of the
flowmeter.
•
Each diagnostics function (e.g. gas bubble detection or electrode deposit detection) can be enabled
individually. After activation, you need to calibrate to match the local conditions, and/or set the limit
values.
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......