10 Error messages
Error no./
area
Text in the LCD display
Cause
Remedy
C190.045
Config.
An alarm is simulated
Switch off alarm simulation
Simulation pulse output is
switched on.
In the "Diagnostics" menu, acti-
vate the simulation mode.
C186.009
Config.
Tx Simulator/Calibrator
mode
Transmitter is operated on the
simulator.
In the "Diagnostics" menu, acti-
vate the simulation mode.
Switch off Calibrator mode
C185.030
Operation
Hold last good knwon value The noise exceeds the bandwidth
set for the noise reduction for a
long period of time.
Switch off noise reduction in the
"Device Setup" menu or contact
the manufacturer.
Switch off noise reduction
Contact service
C184.010
Config.
The Flowrate is set to zero The function of the digital input DI
is set to "Flowrate to Zero" and the
digital input DI is set to high signal
(DC 24 V).
Set the digital input (DI) to low sig-
nal (DC 0 V).
Check digital in terminals
81, 82
C182-008
Config.
Flow simulation
Simulation pulse output is
switched on. One of the following
functions is simulated: flow [%] or
flow [unit] or flow velocity.
These values in simulation mode
do not represent the conditions in
the system.
In the "Diagnostics" menu, switch
off the simulation mode.
Switch off Simulation Mode
C178.000
Config.
Simulated/Fixed Current
Output
The current output is simulated
and is currently set to a specific
path.
In the "Process Alarm" menu,
switch off the simulation mode or
set the HART address to 0 in the
"Communication" menu.
Simulation Mode?
The error message is displayed if
the HART address is not 0 (HART
Multidrop mode, current output
permanently set to 4 mA).
HART address > 0?
C177.015
Config.
HART Addr. <>0
HART address is not 0
(HART Multidrop mode, the cur-
rent output is permanently set to
4 mA.)
Set the HART address to 0 in the
"Communication" menu.
Multidrop mode
Set HART Addr. = 0
C176.011
Config.
Totalizer Stop
The function of the digital input DI
is set to "external totalizer stop"
and the digital input DI is set to
high signal (DC 24 V).
Set the digital input (DI) to low sig-
nal (DC 0 V).
Check digital in terminals
81, 82
C175.013
Config.
Totalizer Reset
The function of the digital input DI
is set to "external totalizer stop"
and the digital input DI is set to
high signal (DC 24 V).
Set the digital input (DI) to low sig-
nal (DC 0 V).
Check digital in terminals
81, 82
C174.002
Config.
Pulse Simulation selected
on DO1
The simulation mode is switched
on
In the "Process Alarm" menu,
switch off the simulation mode.
Switch off Simulation Mode
C172.004
Config.
Pulse Simulation selected
on DO2
The simulation mode is switched
on
In the "Process Alarm" menu,
switch off the simulation mode.
Switch off Simulation Mode
C168.001
Config.
Logic Simulation selected
on DO1
The simulation mode is switched
on
In the "Process Alarm" menu,
switch off the simulation mode.
Switch off Simulation Mode
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......