10 Error messages
162
10.2.2
Function Check
F251.040
Electronics
Self Check Alarm
The SIL monitor function detected
an error in the transmitter.
Replace transmitter or contact
manufacturer
F250.016
Electronics
Tx memory fault detected
Error in the transmitter electronics
unit
Replace electronics or contact
manufacturer
Contact service
F248.036
Sensor
Incompatible snsTx+
Calibration mode is not compati-
ble
Contact manufacturer
sns are not the same series
F246.032
Electronics
Defect digital potentiometer
Transmitter Hardware fault-
Contact service
Internal, digital common mode re-
jection potentiometer defective
Replace electronics or contact
manufacturer
F245.047
Electronics
Stack NV Corrupt
Contact service
The internal stack memory for
PROFIBUS-PA/FOUNDATION
fieldbus is defective.
Replace electronics or contact
manufacturer
F244.031
Electronics
Internal supply voltage er-
ror
Internal voltage supply for the
transmitter is defective
Replace electronics or contact
manufacturer
Contact service
F236.24
Operation
DC to High
Multi-phase measurement media
that produce a very high level of
noise
Check the electrical connections
and ground of the device
Lot of NV-Resets
Refer to instr. Manual
Stones or solids that produce a
very high level of noise
Activate empty pipe detector and
calibrated when the measuring
pipe is empty
Galvanic voltages at the measur-
ing electrodes
Contact manufacturer
Conductivity of fluid is not evenly
distributed in the measurement
medium (e.g. directly after injec-
tion points)
F232.022
Electronics
Driver Error Uref = 0
Incorrectly wired (terminals M1,
M2) or cable break/cable short-cir-
cuit.
Check whether the wiring (termi-
nals M1, M2) is connected correct-
ly, cable break, cable short-circuit
Check wiring for open cir-
cuit
Defective fuse in the coil circuit or
moisture in the terminal box
Check fuse for the coil circuit
Check fuse
Check the terminal box for mois-
ture
F228.020
Electronics
Error in coil circuit
Incorrectly wired (terminals M1,
M2) or cable break/cable short-cir-
cuit.
Check whether the wiring (termi-
nals M1, M2) is connected correct-
ly, cable break, cable short-circuit
Check wiring
for short-circuit
Defective fuse in the coil circuit
Check fuse for the coil circuit
F226.019
Electronics
AD Converter saturated
Signal at the input for the AD con-
verter exceeds the maximum val-
ue of 2.5 V.
If the pipeline is empty, check
whether the empty pipe detection
function is activated.
Chekc empty pipe or
Galv. Voltage
No further measurement is possi-
ble
In the "Diagnostics" menu, switch
on the empty pipe detection.
Check whether the current flow
exceeds the set flow range limit
value. If so, then increase the
measuring range limit value Q
max
.
Error no./
area
Text in the LCD display
Cause
Remedy
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......