133
8 Parameterization
Italics = Parameter can only be viewed in the "Advanced" password level
Menu/parameter
Value range
Description
Communication/
Service Port
Max. Baud Rate
2400, 4800, 9600, 19200,
38400
Transmission speed setting (Baud Rate) for the infra-
red service port
HART Access
ON/OFF
Activation/deactivation of the infrared service port
Menu/parameter
Value range
Description
Communication/
PROFIBUS
The menu is only displayed for devices with PROFI-
BUS-PA.
PA Addr. (-BUS-)
0 to 126
The "Profibus" menu is only displayed if this option
was ordered.
Display of the slave address
Default setting: 126
For further information
see 7.3
Ident Nr. Selector
0x9700, 0x9740, 0x3430
Selection of Ident Nr. Selector
The parameter can only be changed when the cyclic
communication is stopped (Comm State = OFF).
Default setting: 0x3430
Comm State
Offline, Operate, Clear, Stop
Display of the communication status
•
Offline: BUS communication deactivated
•
Operate: cyclic communication running
•
Clear: device is initialized
•
Stop: cyclic communication stopped, BUS com-
munication still active
Al1-Q Flowrate
Display only
Current flow in the set unit from the transducer block
flow including status
Tot1-Q Flowrate
Display only
Current counter reading in the set unit from the trans-
ducer block flow including status
Tot2-Q Flowrate
Display only
Current counter reading in the set unit from the trans-
ducer block flow including status
AI2-Internal Tot Fwd
Display only
Current counter reading of the supply counter in the
set unit from the transducer block spec flow including
status
Al3-Internal Tot Rev
Display only
Current counter reading of the return counter in the
set unit from the transducer block spec flow including
status
Al4-Diagnostics
a
Display only
Current output value including status
The channel can be selected with the "Al4 Channel"
parameter.
To display active values, switch on sensor measure-
ments, and/or conductivity measurements.
a
Al4-Channel
a
Sensor Temp, Conductivity
Selection of the channel output by AI4
The PV_SCALE and OUT_SCALE structure is not
adjusted.
a
Only available for JUMO flowTRANS MAG S02/H02
?
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......