8 Parameterization
140
Italics = Parameter can only be viewed in the "Advanced" password level
Cond. Detection
The conductivity measurement can be used in the nominal width range DN 10 to 300. More informa-
tion
see chapter 9 "Advanced diagnostics functions", page 153.
Menu/parameter
Value range
Description
Diagnostics/Diagnostics Control/
Cond. Detection
a
Cond. Detection
ON/OFF
Activates the "Conductivity Detection" function
Default setting: OFF
One Shot Cond.
The conductivity is measured cyclically at specified
intervals. A one-off measurement can be initiated
here.
Cond. Value
Displays the conductivity
Adj. Cond. Value
5 to 20,000 µS/cm
The conductivity needs to be calibrated for the local
medium. The conductivity measured using a sepa-
rate measuring device can be entered here.
Cond. Min Alarm
5 to 20,000 µS/cm
Sets the minimum limit value for conductivity. When
the value drops below this limit, an alarm is triggered.
Default setting: 5 µS/cm
Cond. Max Alarm
5 to 20,000 µS/cm
Sets the maximum limit value for conductivity. When
the value exceeds this limit, an alarm is triggered.
Default setting: 20,000 µS/cm
Elec. Imp. E1-GND
Display only
Current impedance between electrode E1 and GND
(ground potential)
Elect Imp. E2-GND
Display only
Current impedance between electrode E2 and GND
(ground potential)
Elect. Imp Min Alarm
0 to 20,000 Ω
Sets the minimum limit value for impedance. When
the value drops below this limit, an alarm is triggered.
Default setting: 0 Ω
Elect. Imp Max Alarm
0 to 20,000 Ω
Sets the maximum limit value for impedance. When
the value exceeds this limit, an alarm is triggered.
Default setting: 20,000 Ω
a
Only available for JUMO flowTRANS MAG S02/H02
?
Menu/parameter
Value range
Description
Diagnostics/Diagnostics Control/
Grounding check
a
Grounding Check
Starts the "Grounding check" function
Power spectrum
Display only
Current power spectrum
Amplitude 1 Value
Display only
Displays the four strongest amplitudes in the power
spectrum
Amplitude 2 Value
Display only
Amplitude 3 Value
Display only
Amplitude 4 Value
Display only
a
Only available for JUMO flowTRANS MAG S02/H02
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......