8 Parameterization
Italics = Parameter can only be viewed in the "Advanced" password level
Menu/parameter
Value range
Description
Input/Output/
... Logic Setup
DO1 Signal
No Function, F/R Signal,
Alarm Signal
Dual Range
a
, Batch End
Contact
a
The menu is only displayed if the "Logic/Logic" func-
tion has been selected in the "DO1/DO2 Function"
menu.
This menu is not displayed in the default setting.
•
F/R Signal: the digital output signals the flow di-
rection
•
Alarm Signal: the digital output functions as an
alarm output; the alarm type is set in the "DO1
Alarm Config" menu
•
Dual Range
a
: The digital output is activated when
measuring range 2 (Qmax2) is selected.
•
Batch End Contact
a
: The digital output is activat-
ed when the configured filling quantity is reached.
Default setting: F/R Signal
... DO1 Alarm Config
Selection of "... DO1 Alarm Config" submenu
The menu is only displayed if the "Alarm Signal" func-
tion has been set in the "DO1 Signal" parameter.
DO1 Action
Normally Open, Normally
Closed
Selection of the switching behavior for the digital out-
put
Default setting: Normally Open
DO2 Signal
No Function, F/R Signal,
Alarm Signal
Dual Range
a
, Batch End
Contact
a
See "DO1 Signal" in this table
...
DO2 Alarm Config
Selection of "... DO2 Alarm Config" submenu
The menu is only displayed if the "Alarm Signal" func-
tion has been set in the "DO2 Signal" parameter.
DO2 Action
Normally Open, Normally
Closed
Selection of the switching behavior for the digital out-
put
Default setting: Normally Open
a
Only available for JUMO flowTRANS MAG S02/H02
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......