165
10 Error messages
10.2.3
Operating the device out of spec
Error no./
area
Text in the LCD display
Cause
Remedy
S149.021
Operation
Coil resistor out of limits
Coil resistance too high: coil or coil
circuit fuse is defective or M1/M2
wired incorrectly or cable break or
medium is too hot.
Check wiring, check coil circuit
fuse, contact manufacturer
Check wiring
Contact service
Coil resistance too low: coil is de-
fective or short-circuit in wiring of
M1/M2.
S148.025
Operation
Empty Pipe
The pipeline in the system is emp-
ty.
Fill pipeline
Check Pipe
S146.043
Operation
Gas Bubble Alarm
Gas bubbles were detected in the
medium. The measured value is
above the set switching threshold.
Check the process
S144.033
Operation
Partially filled pipe (TFE)
The partial filling detection is
tripped.
Check the process, fill pipeline
Check pipe or adjust Detec-
tor
S143.042
Operation
Electrode Coating Alarm
Insulating or conductive deposits
were detected on measuring elec-
trodes. The deposit value is above
the set switching threshold.
Check pipe, rinse pipeline, clean
measuring electrodes
S142.041
Operation
Conductivity Alarm
The medium conductivity is out-
side the set limit values.
Check process, adjust alarm limits
if necessary.
S141.046
Operation
Sensor and or Housing
Temperature to high
The sensor temperature is outside
the set limit values.
Check process, adjust alarm limits
if necessary.
S140.007
Operation
Flowrate > 103 %
The flow in the system exceeds
the set measuring range end val-
ue by more than 3 %.
In the "Easy Setup - Q
max
" menu,
increase the measuring range end
value
Check Flowrate
Check Range Setting
S136.006
Operation
Max Alarm
The current flow in the pipeline is
greater than the set max. alarm.
Reduce the flow or increase the
value for the max. alarm.
Flowrate
S132.005
Operation
Min Alarm
The current flow in the pipeline is
greater than the set max. alarm.
Reduce the flow or increase the
value for the max. alarm.
Flowrate
S124.029
Operation
Electr. Impedance too high This may be caused by insulating
deposits on the electrodes or con-
ductivity that is too low or an emp-
ty measuring pipe.
If the pipeline is empty, check
whether the pipe detection func-
tion is switched on.
Coating?
In the "Diagnostics" menu, switch
on the empty pipe detection.
Conductivity?
Check conductivity, check depos-
its on the electrodes.
Empty Pipe?
Increase the value for "Elec. Imp.
Max. Alarm" in the "Diagnostics -
Alarm Limits" menu.
S122.026
Operation
Short circuit E1
Galvanic voltages
Increase the value in the "Diag-
nostics - Alarm Limits - Electr. V
Max Alarm" menu and decrease
the value for "Electr. V Min Alarm".
E2 with shield
S120.023
Operation
Electrode Noise too high
The noise at the measuring elec-
trodes is above the limit value
Check the process
Switch on Noise reduction
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......