141
8 Parameterization
Italics = Parameter can only be viewed in the "Advanced" password level
TFE Detector
To use this function, the sensor must be equipped with a measuring electrode for the partial filling detec-
tion (option).
The installation position must be horizontal with the terminal box facing up. This function is useful for
sensors as of DN 50 without any Ex-protection or with Ex-protection for zone 2. More information
see
chapter 9 "Advanced diagnostics functions", page 153.
Adjust TFE Full
Prior to starting the (manual/automatic) calibration, make sure that:
•
There must be no flow through the sensor (close all valves, shut-off devices, etc.).
•
The sensor must be completely filled with the medium to be measured.
Italics = Parameter can only be viewed in the "Advanced" password level
Menu/parameter
Value range
Description
Diagnostics/Diagnostics Control/
TFE Detector
TFE Detector
Activation of the "Partial filling detector" function
(TFE)
Adjust TFE Full
Partial filling detection must be calibrated to match
the local conditions.
Starts the automatic calibration of partial filling detec-
tion (see note) Cross-reference
TFE Threshold
-
Manual fine adjustment of the switching threshold
The switching threshold is determined during the au-
tomatic calibration.
If the current value is above the set switching thresh-
old, a message is output on the LCD display and the
alarm is activated at the digital output, if configured
accordingly.
TFE Value
Displays the current measured value
?
?
Menu/parameter
Value range
Description
Diagnostics/Diagnostics Control/
SIL Detector
SIL Detection
ON/OFF
The detection level of the safety-relevant parts is in-
creased by switching the detector on.
Default setting: OFF
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......