4 • Troubleshooting
Doc. 065051-03 1/08
107
•
[348] A failure has occurred during power up! The status of all tests
run during power up are indicated on the Diagnostic Tests
screen which follows this error message.
At power-up, the AS electronics are tested. If one or more tests fails,
this message appears. On the
DIAGNOSTIC TESTS
screen, an “F”
indicates that a test failed and a “P” indicates it passed.
Possible Causes:
One or more of these tests has failed, indicating
a problem with that portion of the electronics. See
for a description of each test.
Actions:
•
If the
SYRINGE COMM
test fails, make sure the sample syringe
cable is installed and connected to the correct connector on the
AS rear panel (see
). If the prep syringe is installed,
check its cable connection, also. Rerun the test.
•
If the
SYRINGE COMM
test still fails or if other tests fail, contact
Dionex.
•
[362] Syringe size <= 1 mL require the 1.2 mL needle assy.
Actions:
See
for a description of the sampling
needle assemblies and the syringe size requirements for each.
•
[363] Syringe size > 1 mL require the 8.5 mL needle assy.
Actions:
See
for a description of the sampling
needle assemblies and the syringe size requirements for each.
•
[366] The computed sample injection volume exceeded the needle
assembly volume. The sample injection volume has been
automatically reduced. Please confirm injection volume
parameter.
Actions:
See
“Maximum Sample Loop Sizes and Injection
for the maximum injection volumes
allowed for each sampling needle assembly and syringe size.
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...