5 • Service
Doc. 065051-03 1/08
163
4. Click
Connect to Timebase
on the
Control
menu. Connect to the timebase in
which the AS is configured. When connected, the controls on the Wellness
panel are enabled (see
5. Under
Calibration: leak detector
, click
tray
.
The AS calibrates the sensor and then uploads the new value to Chromeleon.
Chromeleon stores the new calibration value as the current value.
6. To test the leak sensor, click
tray
under
Diagnostic Tests: leak detector
.
Figure 5-30. Dionex Autosampler Wellness Panel