AS Autosampler Operator’s Manual
122
Doc. 065051-03 1/08
•
[475] Cannot run schedule. No DSP detected.
Possible Causes:
Electronic malfunction
Actions:
Turn off the AS power and then turn on the power again. If
the message reappears, contact Dionex.
•
[476] Cannot run schedule. No sample syringe detected.
Possible Causes:
•
Syringe cable not plugged into the rear of the AS
•
Faulty syringe cable or syringe
Actions:
•
Check the syringe cable connection.
•
Replace the cable. Contact Dionex.
•
[479] Non-recoverable error. Both internal and external inject
valves are connected. Please, power down and disconnect one
inject valve before continuing.
Possible Causes:
An injection valve is installed in both the
autosampler compartment and in another module.
Actions:
Disconnect one of the valves.
•
[480] Timed out waiting for tray temperature to stabilize. Either
temperature control is broken, or set temperature is too far
away from ambient.
The sample temperature control option was not able to reach the
temperature set point within the time allotted.
Possible Causes:
•
Ambient temperature too high or low
•
Airflow to cool the Peltier devices is restricted
•
Malfunction of the sample temperature control option
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...