AS Autosampler Operator’s Manual
232
Doc. 065051-03 1/08
C.3 Detail Status Screen
Use the
DETAIL STATUS
screen to:
•
Display detailed information about the status of
AS
operating parameters.
This screen displays all the information shown on the
MAIN STATUS
screen,
plus additional parameters.
•
Set several parameters for Direct Control operation.
NOTE See
for descriptions of the parameters
common to the
MAIN STATUS
and the
DETAIL STATUS
screens.
Figure C-4. Detail Status Screen
Control Mode
SCH
(schedule) or
DIRECT
CONTROL
Press
Select
Δ
or
Select
∇
to select
SCH
(schedule) or
DIRECT
CONTROL
mode. When the AS is in
LOCKED RMT
mode, Direct
Control is always selected.
In Schedule mode, the screen displays the number of the currently
running schedule, the schedule line, and the method. Enter the schedule
number to run (1 to 9) in the
SCH
field.
Screen Field
Description
Help prompt
INJECTING
VIAL#:
TIME:
INJ#:
VOL:
METHOD:
TRAY:
TRAY:
LOOP:
0.0
25
25
25.0
1
1/3
01
P10 mL
uL
uL
min
TTL1
TTL2
RLY1
RLY2
INJ VLV
0
1
0
0
L
NORMAL FULL
SCH 1
LOCAL
DETAIL
LINE 1
20 °C
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...