AS Autosampler Operator’s Manual
200
Doc. 065051-03 1/08
3. Connect the blue sample out line from port 2 on the diverter valve to
port S on the injection valve of system #2.
4. Route the green waste lines connected to port W (6) on each injection
valve through the side slots to the AS drip tray. Insert one line into the
small round opening in the tray (see
). Insert the other
line about 12-mm (1/2-in) into the large drain opening in the tray.
5. Calibrate the inject port volume (the volume of the needle seal and
diverter valve line) for each system. Refer to
instructions.
B.3.7
Connecting a Second AS to a Dual ICS-3000 System
When an ICS-3000 Detector/Chromatography module (DC) is configured
with two injection valves, two AS Autosamplers can be connected to the
DC. The first AS is installed on the left side of the DC and the second AS
is installed on the right side.
A Second AS Kit (P/N 062578) is available for connecting the second AS
to the DC. The kit consists of a 24-in needle seal assembly (P/N 062564)
and a metal tubing chase (062579) and mounting screws (P/N 057058).
The tubing chase routes the needle seal line across the front of the AS
compartment to the DC.
1. Remove the existing needle seal assembly from the AS and install the
24-inch assembly (see
2. After reinstalling the drip tray, do not fully tighten the drip tray screw.
3. Use the two screws to attach the metal tubing chase to the front of the
4. Route the needle seal line behind the face of the drip tray through the
tubing chase to the left side of the AS (see
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...