AS Autosampler Operator’s Manual
278
Doc. 065051-03 1/08
Selecting Direct Control Parameters
1. Go to the
DETAIL
STATUS
screen (press
Menu
and
2
).
2. If the AS is not in Direct Control mode, position the cursor in the control
mode field and press
Select
Δ
or
Select
∇
to select
DIRECT CONTROL
. Press
Enter
.
Figure D-4. Detail Status Screen Example
3. Position the cursor in the field for the parameter to be changed.
4. To set the
TRAY
temperature:
a. If the field displays “--”, temperature control is currently turned off. Press
Select
Δ
or
Select
∇
to turn on temperature control.
b. Press numeric buttons to enter the desired temperature setting.
5. For other parameters, press
Select
Δ
or
Select
∇
to select the desired setting.
•
TTL and relay fields: Select
1
for on or
0
for off.
•
Inject valve(s): Select
L
to set the valve to load or
I
to set the valve to
inject.
6. After setting a parameter, press
Enter
.
NOTE The
TRAY
field is available only when the sample
temperature control option is installed. Tray
temperature can be set to between 4
°
C and 60
°
C.
Dionex does not recommend using the 10 mL plastic tray with the
sample temperature control option because the plastic tray is
inefficient at heating and cooling the vials. In addition, the tray is not
designed for temperatures above 40
°
C (104
°
F) and may become
deformed over time if used above 40
°
C.
Help Prompt
IDLE
VIAL#:
TIME:
INJ#:
VOL:
METHOD:
TRAY:
TRAY:
LOOP:
20 °C
25
100
2 mL
uL
uL
min
TTL1
TTL2
RLY1
RLY2
0
1
0
0
LOCAL
DETAIL
DIRECT CONTROL
INJ VLV L
NORMAL FULL
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...