5. To verify that the syringe drive shaft is in the correct position, fill a syringe
with deionized water and insert it into port D on the valve. Push liquid
through the port and verify that it exits through port C.
6. Each valve port must contain a white Teflon washer. Check the inside of each
port and insert a washer (P/N 061315), if one is not already present. The
washers are supplied with the valve and with the
AS
Preventive Maintenance
Kit (P/N 060581).
7. Holding the syringe at the top fitting, screw it into the bottom of the valve at
port D and tighten fingertight.
8. With valve ports A and D at the 12 o’clock and 6 o’clock positions,
respectively, push the valve onto the syringe drive shaft.
9. Replace the two valve thumbscrews and partially tighten.
10. Carefully align the syringe piston with the connecting screw mount on the
syringe drive.
11. Insert the syringe drive connecting screw and tighten fingertight. Finish
tightening the valve thumb screws.
12. Reconnect the valve liquid lines. Tighten the fittings fingertight.
13. On the
MENU of SCREENS
,
press
5
and
3
to go to the
PLUMBING
CONFIGURATION
screen. Set the initial position (see
14. On the
PLUMBING CONFIGURATION
screen, move the cursor to the
SAMPLE
SYRINGE VOLUME
field. Press
Select
Δ
or
Select
∇
to select the syringe
volume. Press
Enter
.
If a fitting leaks after operation is resumed, tighten it slightly. Do not
overtighten; overtightening can damage the inside of the valve.
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...