D • Front Panel Control
Doc. 065051-03 1/08
291
•
If the source is a vial, the range depends on the volume of the syringe:
•
If you specify a prime or flush volume greater than the syringe
volume and the source is the flush or a reagent reservoir, the syringe
takes multiple strokes to complete the operation.
D.5.4
Selecting Method Setup Parameters
To go to the
METHOD SETUP
screen:
1. Go to the
METHOD MENU.
2. Position the cursor on the
2
field
(SETUP)
.
3. Press
Enter
.
shows the default setup parameters.
Sample Syringe
Volume (
µL)
Reagent Prime/Flush
Range (
µL)
Default (
µL)
100 to 1,000
100 to 1,000
1,000 (any size vial)
5,000
100 to 4,700
1,000 (1.5 mL vial) 4,000
(10 mL vial)
10,000
100 to 8,000
1,000 (1.5 mL vial) 8,000
(10 mL vial)
Figure D-8. Method Setup Screen (Default)
Help Prompt
METHOD:
METHOD SETUP
2
min
mm
---
---
NO
CYCLE TIME:
SAMPLE NEEDLE HEIGHT:
TRAY TEMPERATURE:
WAIT FOR TEMP STABLE:
25
°C